This instrument is used to determine the distillation range of petroleum products in accordance with GB/T 6536, complying with ASTM D86 and IP123. It automatically controls the heating process and distillation rate, and records and prints all test data.
This instrument is used to determine the distillation range of petroleum products in accordance with GB/T 6536, complying with ASTM D86 and IP123. It automatically controls the heating process and distillation rate, and records and prints all test data.
This simulated distillation device comprises an automatic bath/distillation temperature control system, a refrigeration system, an automatic level tracking system, a safety system, and other components. The instrument uses multi-thread operation and control to enable automated operation, control, computation, and display, thereby improving intelligent and automatic measurement. The instrument applies a fuzzy temperature control principle. A Freon compressor is used in the refrigeration equipment for temperature control, enabling precise regulation of the condenser and receiving chamber temperatures. The temperature measurement system adopts high-precision thermal resistance elements for accurate steam temperature measurement. The instrument also uses an imported high-precision level tracking system to precisely measure distillation volume with an accuracy of 0.1 ml.
To facilitate human-machine interaction, the system features a true-color touch screen. Users can set parameters via the touch screen for real-time monitoring of operating parameters, recording of critical temperatures, tracing of temperature–volume curves, storage of 256 test data groups, and retrieval of historical data for various oils.
This instrument complies with GB/T 6536-2010. The user can enable/disable automatic pressure calibration. The system includes a high-accuracy atmospheric pressure measuring device. In addition, the instrument is equipped with temperature and pressure monitoring, auxiliary equipment, a fire extinguisher, and level tracking equipment for automatic monitoring. In the event of a malfunction, the system will automatically prompt for immediate action to prevent accidents.
1. Compact, attractive design, and easy to operate.
2. Fuzzy temperature control, high precision, and fast response.
3. 10.4" large color touch screen for easy use.
4. High level tracking accuracy.
5. Automatic distillation process and monitoring.
Power |
AC 220V ±10%, 50 Hz |
|||
Heating power |
2 kW |
|||
Cooling power |
0.5 kW |
|||
Steam temperature |
0–400 °C |
|||
Oven temperature |
0–500 °C |
|||
Refrigeration temperature |
0–60 °C |
|||
Refrigeration accuracy |
±1 °C |
|||
Temperature measurement accuracy |
±0.1 °C |
|||
Volume accuracy |
±0.1 ml |
|||
Fire alarm |
Extinguish by nitrogen (prepared by customer) |
|||
Sample state |
suitable for natural gasoline (stable light hydrocarbons), motor gasoline, aviation gasoline, jet fuel, special boiling point solvent, naphtha, mineral spirits, kerosene, diesel fuel, gas oil, and distillate fuels. |
|||
Indoor working environment |
Temperature |
10–38 °C (reference: 10–28 °C) |
Humidity |
≤70% |