The interfacial tensiometer is applicable to GB/T 6541. It determines interfacial tension or surface tension based on the molecular forces between phases, reflecting the physical and chemical properties of the liquid and its composition; it is used as an important product-quality inspection indicator in related industries. PS-ZL203 uses the ring method (platinum ring method) to measure the surface tension of various liquids (liquid–gas interface) and the interfacial tension of liquids (liquid–li
The interfacial tensiometer is applicable to GB/T 6541. It determines interfacial tension or surface tension based on the molecular forces between phases, reflecting the physical and chemical properties of the liquid and its composition; it is used as an important product-quality inspection indicator in related industries. PS-ZL203 uses the ring method (platinum ring method) to measure the surface tension of various liquids (liquid–gas interface) and the interfacial tension of liquids (liquid–li
The original fast-response electromagnetic force balance sensor is used to improve side precision and linearity. The instrument calibration is performed at a single reference point only, addressing the limitation of previous sensor generations that required multi-point calibration. It also eliminates the need for a zeroing potentiometer and a full-range potentiometer. Equivalent tension and load are displayed in real time. An integrated temperature detection circuit provides automatic temperature compensation for measurement results. The 240 × 128 dot-matrix LCD features no identification key and includes a screen protection function. A time-stamped history record stores up to 255 data sets. The built-in high-speed thermosensitive printer offers fast, clear printing with offline printing capability.
|
Name |
Indicator |
|
Measuring range |
0–200 mN |
|
Accuracy |
0.1% of reading ± 0.1 mN/m |
|
Sensitivity |
0.1 mN/m |
|
Resolving power |
0.1 mN/m |
|
Supply voltage |
AC 220 V ± 10% |
|
Power frequency |
50 Hz ± 2% |
|
Power |
≤ 20 W |
|
Applicable temperature |
10–40 ℃ |
|
Applicable humidity |
< 85% RH |
|
Dimensions (W × H × D) |
200 mm × 330 mm × 300 mm |
|
Net weight |
Approx. 5 kg |